Compare products
Manufacturer
Meril
SIS Medical
Meril
Model
MeRes100
OPN NC
BioMime
Category
Coronary Intervention
Coronary Intervention
Coronary Intervention
Purchase
By quotation
By quotation
By quotation
Warranty
Not listed
Not listed
Not listed
Scaffold material
Bioresorbable PLLA (poly L-lactide)
Not listed
Not listed
Strut thickness
100 µm
Not listed
65 µm
Drug
Sirolimus, 1.25 µg/mm²
Not listed
Sirolimus, 1.25 µg/mm²
Polymer
Bioresorbable PDLLA (poly D,L-lactide)
Not listed
PLLA and PLGA (biodegradable + biocompatible), 2 µm thickness
Architecture
Hybrid design: closed cells at ends, open cells in middle
Not listed
Not listed
Diameters
2.50, 2.75, 3.00, 3.25, 3.50, 4.00, 4.50 mm
Not listed
2.00, 2.25, 2.50, 2.75, 3.00, 3.50, 4.00, 4.50 mm
Lengths
13, 16, 19, 24, 29, 32, 37, 40 mm
Not listed
8, 13, 16, 19, 24, 29, 32, 37, 40, 44, 48 mm (8 mm not in 2.00 mm diameter)
Delivery system
Rapid exchange
Not listed
Rapid exchange; 0.016" tip entry profile
Type
Not listed
Super-high pressure non-compliant PTCA dilatation catheter, twin-wall (balloon-in-balloon) design
Not listed
Balloon diameters
Not listed
1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5 mm
Not listed
Balloon lengths
Not listed
10, 15, 20 mm
Not listed
Nominal pressure
Not listed
10 atm
Not listed
Rated burst pressure
Not listed
35 atm
Not listed
Minimum guiding catheter
Not listed
6F (1.5–3.5 mm); 7F (4.0–4.5 mm)
Not listed
Guidewire
Not listed
0.014" (0.36 mm)
Not listed
Usable catheter length
Not listed
140 cm
Not listed
Coating / markers
Not listed
Hydrophobic coating; dual Pt/Ir markers
Not listed
Required inflation device
Not listed
High-pressure inflation device rated to at least 35 atm
Not listed
Regulatory
Not listed
CE marked; FDA cleared
Not listed
Stent material
Not listed
Not listed
Cobalt chromium L605
Stent architecture
Not listed
Not listed
Hybrid cell design (closed cells at ends, open cells in middle) with non-linear links
Balloon overhang
Not listed
Not listed
Less than 0.5 mm