Skip to content

Compare products

Aspiration catheter

Aspiration catheter

Aspiration catheter from Meril's vascular intervention range.

Learn more
OPN NC catheter hub

OPN NC

Super high-pressure PTCA balloon, 35 atm rated burst pressure.

Learn more
BioMime sirolimus-eluting coronary stent system

BioMime sirolimus-eluting coronary stent system

Sirolimus-eluting coronary stent system with 65 µm struts.

Learn more

Manufacturer

Meril

SIS Medical

Meril

Model

Not listed

OPN NC

BioMime

Category

Coronary Intervention

Coronary Intervention

Coronary Intervention

Purchase

By quotation

By quotation

By quotation

Warranty

Not listed

Not listed

Not listed

Product name

Aspiron

Not listed

Not listed

Indication

Removal of fresh, soft emboli and thrombi from coronary and peripheral vessels

Not listed

Not listed

Catheter design

Dual lumen, rapid exchange

Not listed

Not listed

Working length

140 cm

Not listed

Not listed

Guide catheter compatibility

6F

Not listed

Not listed

Guidewire compatibility

0.014" (0.36 mm)

Not listed

Not listed

Coating

Hydrophilic, on the distal shaft

Not listed

Not listed

Accessories

30 mL suction syringes, one-way stopcock, extension tube, flushing needle, 70 µm filter caps

Not listed

Not listed

Regulatory

FDA 510(k) K190619 (Class II)

CE marked; FDA cleared

Not listed

Biocompatibility

Tested to ISO 10993-1

Not listed

Not listed

Type

Not listed

Super-high pressure non-compliant PTCA dilatation catheter, twin-wall (balloon-in-balloon) design

Not listed

Balloon diameters

Not listed

1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5 mm

Not listed

Balloon lengths

Not listed

10, 15, 20 mm

Not listed

Nominal pressure

Not listed

10 atm

Not listed

Rated burst pressure

Not listed

35 atm

Not listed

Minimum guiding catheter

Not listed

6F (1.5–3.5 mm); 7F (4.0–4.5 mm)

Not listed

Guidewire

Not listed

0.014" (0.36 mm)

Not listed

Usable catheter length

Not listed

140 cm

Not listed

Coating / markers

Not listed

Hydrophobic coating; dual Pt/Ir markers

Not listed

Required inflation device

Not listed

High-pressure inflation device rated to at least 35 atm

Not listed

Drug

Not listed

Not listed

Sirolimus, 1.25 µg/mm²

Polymer

Not listed

Not listed

PLLA and PLGA (biodegradable + biocompatible), 2 µm thickness

Stent material

Not listed

Not listed

Cobalt chromium L605

Strut thickness

Not listed

Not listed

65 µm

Stent architecture

Not listed

Not listed

Hybrid cell design (closed cells at ends, open cells in middle) with non-linear links

Diameters

Not listed

Not listed

2.00, 2.25, 2.50, 2.75, 3.00, 3.50, 4.00, 4.50 mm

Lengths

Not listed

Not listed

8, 13, 16, 19, 24, 29, 32, 37, 40, 44, 48 mm (8 mm not in 2.00 mm diameter)

Delivery system

Not listed

Not listed

Rapid exchange; 0.016" tip entry profile

Balloon overhang

Not listed

Not listed

Less than 0.5 mm